TRULIANT
Medical device: Uncoated knee femur prosthesis, metallic. 2 companies paid 49 providers $807K in 2025 payments naming it.
- Medical device
- Paid for by Exactech, Inc., Advita Ortho, Llc
- FDA Class 2
Companies paying for it (2025)
- EXACTECH, INC.$806K
- ADVITA ORTHO, LLC$1,389
Safety record
- FDA recalls naming it6
- Device problem reports1,326
- FDA risk classClass 2 (moderate)
Reports to FDA don't prove the product caused the event.
About this device
An implantable device designed to replace the femoral condyles (femoral component) during primary bicondylar replacement of the knee joint, femoral bone resection, or to replace a dysfunctional knee prosthesis (revision); it is made of metal [e.g., cobalt-chrome (Co-Cr), cobalt-chrome-molybdenum (Co-Cr-Mo), titanium (Ti)] and is not coated with a material intended to improve fixation and stability by promoting bone ingrowth. The device is designed to articulate with tibial and, if required, patellar prosthetic components of the knee joint and its implantation is intended to be performed with …
- Brand
- Truliant
- Company
- Exactech, Inc.
- Device type
- Uncoated knee femur prosthesis, metallic
- FDA product code
- JWH · Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
- FDA class
- Class 2 (moderate risk)
- Specialty
- Orthopedic
- Sold
- by prescription
- Implant
- Yes
- Status
- In Commercial Distribution
- GUDID record
- 10885862314659
FDA clearance or approval
| Number | Kind | Device | Applicant | Decided |
|---|---|---|---|---|
| K170240 | 510(k) | Truliant Femoral Components | Exactech, Inc. | 2017-02-23 |
A 510(k) clearance means FDA found the device much like one already sold. A PMA approval means FDA reviewed clinical evidence, which it requires for high-risk devices.
FDA's device database (GUDID) record of the device identifier named most in the payments: one model of the product line.
FDA recalls
6 recalls naming it since 2004 (Class I, the most serious: 0; Class II: 6; Class III: 0).
| Started | Recall | Class | Firm | Product | Reason |
|---|---|---|---|---|---|
| 2023-11-21 | Z-0586-2024 | Class II | Exactech, Inc. | Exactech Truliant, Activit-E Knee Insert, Catalog Numbers: a) 02-023-02-0029, b) 02-023-02-0032, c) 02-023-02-0035, d) 02-023-02-0038, e) 02-024-35-1512, f) 02-024-35-1515, g) 02-024-35-1519, h) 02-024-35-2010, … | Exactech has received 3 complaints for 4 devices related to a loss of vacuum in the inner-most vacuum bag. |
| 2023-11-21 | Z-0585-2024 | Class II | Exactech, Inc. | Exactech Truliant Knee System, Catalog Numbers: a) 02-012-35-3509, b) 02-022-35-1509, c) 02-022-35-1513, d) 02-022-35-3010, e) 02-022-35-3509, f) 02-022-35-3510, g) 02-022-35-3511, h) 02-022-35-4511, i) 02-022-44-3512, … | Exactech has received 3 complaints for 4 devices related to a loss of vacuum in the inner-most vacuum bag. |
| 2021-08-30 | Z-0023-2022 | Class II | Exactech, Inc. | TRULIANT Knee System including TRULIANT Tibial Inserts, also labeled as the following: a. TRULIANT CR Tibial Inserts; b. TRULIANT CR Slope + Tibial Inserts; c. TRULIANT CR Slope ++ Tibial Inserts; d. TRULIANT CRC Tibial … | Inserts were packaged in vacuum bags that lacked an additional oxygen barrier layer. |
| 2020-01-27 | Z-1364-2020 | Class II | Exactech, Inc. | Truliant Splined Stem Extension, 18mm x 120mm - Product Usage: The devices are used in a cemented total knee prosthesis system, which provides femoral components, tibial inserts, stems, augments, and surgical instrumentation for use in … | These Truliant Splined Stem Extension, 18mm x 120mm were found to have a labeling error. The upper right corner of the label incorrectly identifies the implant size as 17mm x 80mm when the implant size is 18mm x 120mm. |
| 2019-03-13 | Z-1208-2019 | Class II | Exactech, Inc. | Truliant Tibial Fit Tray Cemented SZ 3F / 3T, Catalog Number 02-022-45-3030 Product Usage: Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented The OPTETRAK Comprehensive Knee Systems are indicated for use in … | 2.5 Truliant Fit Trays were mismarked as 3 |
| 2017-06-23 | Z-3018-2017 | Class II | Exactech, Inc. | Truliant Tibial Trial Handle. To assist the surgeon in the implantation of Truliant Knee system components according to a conventional technique for total knee replacement. Specifically, these instruments are reusable, surgically … | The Truliant Tibial Trial Handle's pin may disassociate from the main body. |
From FDA enforcement reports that name this device, by a company paying for it. Class I is the most serious recall and Class III the least.
The full lists, detailed tables and downloads are in the Data view.