TRIGEN
Medical device: Femur intramedullary nail. 1 company paid 201 providers $1.30M in 2025 payments naming it.
- Medical device
- Paid for by Smith+Nephew, Inc.
- FDA Class 2
Companies paying for it (2025)
- SMITH+NEPHEW, INC.$1.30M
Safety record
- FDA recalls naming it13
- Device problem reports973
- FDA risk classClass 2 (moderate)
Reports to FDA don't prove the product caused the event.
About this device
A rigid rod made of metal or other material intended to be implanted into the intramedullary canal of the femur to immobilize the ends of a fractured bone in position to promote healing; it is not dedicated to extension beyond the joint to an adjacent bone (i.e., not dedicated to arthrodesis/joint fusion). It may be a locked or unlocked model and may include components to assist fixation of more proximal or distal conditions. This is a single-use device.
- Brand
- TRIGEN
- Company
- Smith & Nephew, Inc.
- Model described
- TRIGEN META RETROGRADE FEMORAL NAIL 11.5MM X 44CM (71653244)
- Device type
- Femur intramedullary nail
- FDA product code
- JDS · NAIL, FIXATION, BONE
- FDA class
- Class 2 (moderate risk)
- Specialty
- Orthopedic
- Sold
- by prescription
- Implant
- Yes
- Status
- In Commercial Distribution
- GUDID record
- 03596010548474
FDA's device database (GUDID) record of the device identifier named most in the payments: one model of the product line.
FDA recalls
13 recalls naming it since 2004 (Class I, the most serious: 0; Class II: 13; Class III: 0).
| Started | Recall | Class | Firm | Product | Reason |
|---|---|---|---|---|---|
| 2026-03-16 | Z-2456-2026 | Class II | Smith & Nephew, Inc. | TRIGEN PROXIMAL STRAIGHT HUMERAL NAIL 8/7 X 16, REF: 71760816 | Humeral nail due to a manufacturing error is missing distal holes and if implanted may lead to revision surgery but, if the missing holes are identified before implantation, an equivalent backup can be used, or a bent nail of the same size, … |
| 2023-07-24 | Z-2444-2023 | Class II | Smith & Nephew, Inc. | TRIGEN TROCHANTERIC ANTEGRADE NAIL LIME 11.5MM X 40CM 130 DEGREE LEFT, REF 71647340, Femur nail | A complaint was received indicating that a package contained a TRIGEN TROCHANTERIC ANTEGRADE NAIL LIME 11.5MM X 40CM 130 DEGREE LEFT implant instead of a TRIGEN TROCHANTERIC ANTEGRADE NAIL LIME 11.5MM X 36CM 130 DEGREE LEFT as described on … |
| 2023-07-24 | Z-2443-2023 | Class II | Smith & Nephew, Inc. | TRIGEN TROCHANTERIC ANTEGRADE NAIL LIME 11.5MM X 36CM 130 DEGREE LEFT, REF 71647336, Femur nail | A complaint was received indicating that a package contained a TRIGEN TROCHANTERIC ANTEGRADE NAIL LIME 11.5MM X 40CM 130 DEGREE LEFT implant instead of a TRIGEN TROCHANTERIC ANTEGRADE NAIL LIME 11.5MM X 36CM 130 DEGREE LEFT as described on … |
| 2020-11-02 | Z-0772-2021 | Class II | Smith & Nephew, Inc. | Bone fixation nail. TRIGEN INTERTAN - Product Usage: indicated for fractures of the femur including simple shaft fractures, comminuted shaft fractures, spiral shaft fractures, long oblique shaft fractures and segmental shaft fracture; | Manufacturing packaging error. |
| 2020-06-24 | Z-2685-2020 | Class II | Smith & Nephew, Inc. | TRIGEN, Intramedullary nail - Product Usage: is intended to provide resistance for screw back-out. It provides additional friction to help hold the screws in place. | A manufacturing error which resulted in the re-sterilization of the ultra-high-molecular-weight polyethylene (UHMWPE) sleeve. |
| 2019-10-07 | Z-0351-2020 | Class II | Smith & Nephew, Inc. | LAG SCREW 3.2MM GUIDE PIN SLEEVE, REF 71674032 - Product Usage: The TRIGEN InterTAN nails are indicated for fractures of the femur including: simple shaft fractures, comminuted shaft fractures, spiral shaft fractures, long oblique shaft … | LAG SCREW 3.2 GUIDE PIN SLEEVE are being recalled due to a manufacturing error. The detent balls in the guide pin sleeve were not drilled to the correct depth, which will cause them to not fully compress. |
| 2017-10-24 | Z-0097-2018 | Class II | Smith & Nephew, Inc. | smith&nephew TRIGEN INTERTAN 10S, REF 71675384, 10 MM X 18 CM NAIL, 130o (degrees), SET SCREW PREASSEMBLED IN NAIL | A single lot of INTERTAN 10S 10 MM X 18 CM 130 D was manufactured with the screw inserted upside down. |
| 2014-07-02 | Z-2258-2014 | Class II | Smith & Nephew, Inc. | TRIGEN(R) INTERTAN(TM) 10MM X 18 CM NAIL, 125o, REF 71675201, QTY (1), STERILE R, Smith & Nephew, Memphis, Inc., TN 38116 TRIGEN(R) INTERTAN(TM) 11.5MM X 18 CM NAIL, 125o, REF 71675202, QTY (1), STERILE R, Smith & Nephew, Memphis, Inc., … | There was an error in manufacturing which resulted in the proximal lag screw hole being positioned incorrectly in several batches of TRIGEN(R); INTERTAN(TM); NAILS. |
| 2014-04-04 | Z-1637-2014 | Class II | Smith & Nephew Inc | TRIGEN INTERTAN 1.5 NAIL: (1) REF 71676515, 10 MM X 10 CM, 125 degree, LEFT, (2) REF 71676556, 11.5 MM X 36 CM, 125 degree, RIGHT, (3) REF 71676558, 11.5 MM X 38 CM, 125 degree, RIGHT, (4) REF 71676561, 11.5 MM X 42 CM, 125 degree, … | A manufacturing error resulted in the set screw being inserted too deep in the affected nails. |
| 2014-04-04 | Z-1636-2014 | Class II | Smith & Nephew Inc | TRIGEN INTERTAN NAIL: (1) REF 71675201, 10 MM X 18 CM, 125 degree, (2) REF 71675202, 11.5 MM X 18 CM, 125 degree, (3) REF 71675203, 13 MM X 18 CM, 125 degree, (4) REF 71675204, 10 MM X 20 CM, 125 degree, (5) REF 71675205, 11.5 MM … | A manufacturing error resulted in the set screw being inserted too deep in the affected nails. |
| 2013-03-06 | Z-1132-2013 | Class II | Smith & Nephew Inc | TRIGEN (TM) L-P SCREW, 4.5 MM X 37.5 MM, TI-6AL-4V, QTY: (1), STERILE R, REF 71645437, Smith & Nephew, Inc. Memphis, TN 38116 USA orthopedic surgery | One batch of 4.5 mm diameter Trigen Low-Profile Bone Screws were incorrectly anodized with a gold coating (indicating a 5 mm diameter) instead of a gray coating (indicating a 4.5 mm diameter). |
| 2013-02-07 | Z-0944-2013 | Class II | Smith & Nephew Inc | INTERNAL HEX RECON SCREW, TI-6AL-4V, REF 7162375, TRIGEN (R) INTERNAL HEX CAPTURED RECON SCREW, QTY: (1), 6.4 MM, 75 MM, TITANIUM, Smith & Nephew, Inc. Orthopedic. Bone screw for intramedullary rod fixation. | 75 mm screws were packaged and labeled using an 80 mm screw outer box and label, while 80 mm screws were packaged and labeled using a 75 mm screw outer box and label. |
| 2013-02-07 | Z-0943-2013 | Class II | Smith & Nephew Inc | INTERNAL HEX RECON SCREW, TI-6AL-4V, REF 7162380, TRIGEN (R) INTERNAL HEX CAPTURED RECON SCREW, QTY: (1), 6.4 MM, 80 MM, TITANIUM, Smith & Nephew, Inc. Orthopedic. Bone screw for intramedullary rod fixation. | 75 mm screws were packaged and labeled using an 80 mm screw outer box and label, while 80 mm screws were packaged and labeled using a 75 mm screw outer box and label. |
From FDA enforcement reports that name this device, by a company paying for it. Class I is the most serious recall and Class III the least.
The full lists, detailed tables and downloads are in the Data view.