TRIGEN

Medical device: Femur intramedullary nail. 1 company paid 201 providers $1.30M in 2025 payments naming it.

  • Medical device
  • Paid for by Smith+Nephew, Inc.
  • FDA Class 2
Paid to providers, 2025 $1.30M 483 payments to 201 providers
Device problem reports 973 0 deaths, 879 injuries (FDA MAUDE)
Companies paying for it 1 in 2025 payments

Companies paying for it (2025)

Safety record

  • FDA recalls naming it13
  • Device problem reports973
  • FDA risk classClass 2 (moderate)

Reports to FDA don't prove the product caused the event.

About this device

A rigid rod made of metal or other material intended to be implanted into the intramedullary canal of the femur to immobilize the ends of a fractured bone in position to promote healing; it is not dedicated to extension beyond the joint to an adjacent bone (i.e., not dedicated to arthrodesis/joint fusion). It may be a locked or unlocked model and may include components to assist fixation of more proximal or distal conditions. This is a single-use device.

Brand
TRIGEN
Company
Smith & Nephew, Inc.
Model described
TRIGEN META RETROGRADE FEMORAL NAIL 11.5MM X 44CM (71653244)
Device type
Femur intramedullary nail
FDA product code
JDS · NAIL, FIXATION, BONE
FDA class
Class 2 (moderate risk)
Specialty
Orthopedic
Sold
by prescription
Implant
Yes
Status
In Commercial Distribution
GUDID record
03596010548474

FDA's device database (GUDID) record of the device identifier named most in the payments: one model of the product line.

FDA recalls

13 recalls naming it since 2004 (Class I, the most serious: 0; Class II: 13; Class III: 0).

The newest FDA recalls naming the product
StartedRecallClassFirmProductReason
2026-03-16Z-2456-2026Class IISmith & Nephew, Inc. TRIGEN PROXIMAL STRAIGHT HUMERAL NAIL 8/7 X 16, REF: 71760816Humeral nail due to a manufacturing error is missing distal holes and if implanted may lead to revision surgery but, if the missing holes are identified before implantation, an equivalent backup can be used, or a bent nail of the same size, …
2023-07-24Z-2444-2023Class IISmith & Nephew, Inc. TRIGEN TROCHANTERIC ANTEGRADE NAIL LIME 11.5MM X 40CM 130 DEGREE LEFT, REF 71647340, Femur nailA complaint was received indicating that a package contained a TRIGEN TROCHANTERIC ANTEGRADE NAIL LIME 11.5MM X 40CM 130 DEGREE LEFT implant instead of a TRIGEN TROCHANTERIC ANTEGRADE NAIL LIME 11.5MM X 36CM 130 DEGREE LEFT as described on …
2023-07-24Z-2443-2023Class IISmith & Nephew, Inc. TRIGEN TROCHANTERIC ANTEGRADE NAIL LIME 11.5MM X 36CM 130 DEGREE LEFT, REF 71647336, Femur nailA complaint was received indicating that a package contained a TRIGEN TROCHANTERIC ANTEGRADE NAIL LIME 11.5MM X 40CM 130 DEGREE LEFT implant instead of a TRIGEN TROCHANTERIC ANTEGRADE NAIL LIME 11.5MM X 36CM 130 DEGREE LEFT as described on …
2020-11-02Z-0772-2021Class IISmith & Nephew, Inc. Bone fixation nail. TRIGEN INTERTAN - Product Usage: indicated for fractures of the femur including simple shaft fractures, comminuted shaft fractures, spiral shaft fractures, long oblique shaft fractures and segmental shaft fracture;Manufacturing packaging error.
2020-06-24Z-2685-2020Class IISmith & Nephew, Inc. TRIGEN, Intramedullary nail - Product Usage: is intended to provide resistance for screw back-out. It provides additional friction to help hold the screws in place.A manufacturing error which resulted in the re-sterilization of the ultra-high-molecular-weight polyethylene (UHMWPE) sleeve.
2019-10-07Z-0351-2020Class IISmith & Nephew, Inc. LAG SCREW 3.2MM GUIDE PIN SLEEVE, REF 71674032 - Product Usage: The TRIGEN InterTAN nails are indicated for fractures of the femur including: simple shaft fractures, comminuted shaft fractures, spiral shaft fractures, long oblique shaft …LAG SCREW 3.2 GUIDE PIN SLEEVE are being recalled due to a manufacturing error. The detent balls in the guide pin sleeve were not drilled to the correct depth, which will cause them to not fully compress.
2017-10-24Z-0097-2018Class IISmith & Nephew, Inc. smith&nephew TRIGEN INTERTAN 10S, REF 71675384, 10 MM X 18 CM NAIL, 130o (degrees), SET SCREW PREASSEMBLED IN NAILA single lot of INTERTAN 10S 10 MM X 18 CM 130 D was manufactured with the screw inserted upside down.
2014-07-02Z-2258-2014Class IISmith & Nephew, Inc. TRIGEN(R) INTERTAN(TM) 10MM X 18 CM NAIL, 125o, REF 71675201, QTY (1), STERILE R, Smith & Nephew, Memphis, Inc., TN 38116 TRIGEN(R) INTERTAN(TM) 11.5MM X 18 CM NAIL, 125o, REF 71675202, QTY (1), STERILE R, Smith & Nephew, Memphis, Inc., …There was an error in manufacturing which resulted in the proximal lag screw hole being positioned incorrectly in several batches of TRIGEN(R); INTERTAN(TM); NAILS.
2014-04-04Z-1637-2014Class IISmith & Nephew Inc TRIGEN INTERTAN 1.5 NAIL: (1) REF 71676515, 10 MM X 10 CM, 125 degree, LEFT, (2) REF 71676556, 11.5 MM X 36 CM, 125 degree, RIGHT, (3) REF 71676558, 11.5 MM X 38 CM, 125 degree, RIGHT, (4) REF 71676561, 11.5 MM X 42 CM, 125 degree, …A manufacturing error resulted in the set screw being inserted too deep in the affected nails.
2014-04-04Z-1636-2014Class IISmith & Nephew Inc TRIGEN INTERTAN NAIL: (1) REF 71675201, 10 MM X 18 CM, 125 degree, (2) REF 71675202, 11.5 MM X 18 CM, 125 degree, (3) REF 71675203, 13 MM X 18 CM, 125 degree, (4) REF 71675204, 10 MM X 20 CM, 125 degree, (5) REF 71675205, 11.5 MM …A manufacturing error resulted in the set screw being inserted too deep in the affected nails.
2013-03-06Z-1132-2013Class IISmith & Nephew Inc TRIGEN (TM) L-P SCREW, 4.5 MM X 37.5 MM, TI-6AL-4V, QTY: (1), STERILE R, REF 71645437, Smith & Nephew, Inc. Memphis, TN 38116 USA orthopedic surgeryOne batch of 4.5 mm diameter Trigen Low-Profile Bone Screws were incorrectly anodized with a gold coating (indicating a 5 mm diameter) instead of a gray coating (indicating a 4.5 mm diameter).
2013-02-07Z-0944-2013Class IISmith & Nephew Inc INTERNAL HEX RECON SCREW, TI-6AL-4V, REF 7162375, TRIGEN (R) INTERNAL HEX CAPTURED RECON SCREW, QTY: (1), 6.4 MM, 75 MM, TITANIUM, Smith & Nephew, Inc. Orthopedic. Bone screw for intramedullary rod fixation.75 mm screws were packaged and labeled using an 80 mm screw outer box and label, while 80 mm screws were packaged and labeled using a 75 mm screw outer box and label.
2013-02-07Z-0943-2013Class IISmith & Nephew Inc INTERNAL HEX RECON SCREW, TI-6AL-4V, REF 7162380, TRIGEN (R) INTERNAL HEX CAPTURED RECON SCREW, QTY: (1), 6.4 MM, 80 MM, TITANIUM, Smith & Nephew, Inc. Orthopedic. Bone screw for intramedullary rod fixation.75 mm screws were packaged and labeled using an 80 mm screw outer box and label, while 80 mm screws were packaged and labeled using a 75 mm screw outer box and label.

From FDA enforcement reports that name this device, by a company paying for it. Class I is the most serious recall and Class III the least.

The full lists, detailed tables and downloads are in the Data view.