SIGMA
Medical device: Tibial insert. 3 companies paid 8 providers $2.84M in 2025 payments naming it.
- Medical device
- Paid for by Depuy Synthes Products, Inc., Medical Device Business Services, Inc. +1
- FDA Class 2
Companies paying for it (2025)
Safety record
- FDA recalls naming it7
- Device problem reports24,330
- FDA risk classClass 2 (moderate)
Reports to FDA don't prove the product caused the event.
About this device
An implantable device designed to articulate the femoral and tibial components of a knee joint prosthesis during primary or revision bicondylar knee joint replacement. It is typically a one-piece contoured polyethylene (PE) insert, or a more complex device (e.g., a guide arm with a stem that is inserted into the tibial tray and a polymer glide) which will better reproduce the flexion of the knee joint due to its capacity for lateral rotation and anterior/posterior movement. This device is typically made of metal and/or a polymer.
- Brand
- SIGMA
- Company
- DEPUY (IRELAND)
- Model described
- SIGMA TIBIAL INSERT FIXED BEARING CURVED PLUS 6 20mm XLK (97-0485)
- Device type
- Tibial insert
- FDA product code
- JWH · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER
- FDA class
- Class 2 (moderate risk)
- Specialty
- Orthopedic
- Sold
- by prescription
- Implant
- Yes
- Status
- In Commercial Distribution
- GUDID record
- 10603295250777
FDA clearance or approval
| Number | Kind | Device | Applicant | Decided |
|---|---|---|---|---|
| K241000 | 510(k) | ATTUNE Revision Knee System; DePuy Knee Prosthesis System Universal Stem Extensions and Universal Femoral Metaphyseal Sleeves; DePuy Sigma PS Femoral Components; DePuy Sigma Cruciate Retaining (C/R) Porocoat Femoral Components; S-ROM NOILES Rotating Hinge Knee System; DePuy P.F.C. SIGMA Total Knee System; DePuy SIGMA Total Knee System; | Depuy Ireland UC | 2024-07-11 |
| K040166 | 510(k) | DEPUY SIGMA XLK TIBIAL INSERTS | DePuy Orthopaedics, Inc. | 2004-02-23 |
A 510(k) clearance means FDA found the device much like one already sold. A PMA approval means FDA reviewed clinical evidence, which it requires for high-risk devices.
FDA's device database (GUDID) record of the device identifier named most in the payments: one model of the product line.
FDA recalls
7 recalls naming it since 2004 (Class I, the most serious: 0; Class II: 7; Class III: 0).
| Started | Recall | Class | Firm | Product | Reason |
|---|---|---|---|---|---|
| 2021-02-22 | Z-1315-2021 | Class II | DePuy Orthopaedics, Inc. | The LCS COMPLETE P.F.C.TM SIGMA RP Mobile Bearing Total Knee System - Product Usage: consists of three components intended to replace the femoral, tibial, and patellar articular surfaces of the knee joint. | Globally the company is removing 37 lots affected lots for MBT & ATTUNE (S+) Tibial Bases due to potential damaged packaging. These implant components are utilized in Total Knee Arthroplasty. The affected tibial bases are designed to be … |
| 2021-01-08 | Z-1103-2021 | Class II | DePuy Orthopaedics, Inc. | Universal Femoral Sleeves - Product Usage: used primarily in revision surgeries with the S-ROM Noiles, LPS, SIGMA, and LCS Knee products. | The taper dimensions may be out of specification due to a production issue during the manufacturing of these lots. |
| 2018-07-18 | Z-2949-2018 | Class II | DePuy Orthopaedics, Inc. | P.F.C. SIGMA Revision Knee System Distal Augment, Model Number 960830 | This unit may be missing the screw/collet Assembly |
| 2017-12-21 | Z-1219-2018 | Class II | DePuy Orthopaedics, Inc. | (1) Sigma HP Cemented Trochlea Size 1 Narrow Left, Catalog Number: 102403100, GTIN: 10603295001614; (2) Sigma HP Cemented Trochlea Size 2 Narrow Left, Catalog Number: 102403200, GTIN: 10603295001621; (3) Sigma HP Cemented Trochlea Size 3 … | The SIGMA¿ HP PFJ Cemented Trochlear Implants, a standalone component of the partial knee system is being recalled based on elevated revision rates observed as part of the company s post market surveillance process. |
| 2016-01-08 | Z-0787-2016 | Class II | DePuy Orthopaedics, Inc. | SIGMA HP MBT Non-Keel Punch Knee Instrument. Designed to be used as an option in stabilizing the tibial trial during trial reduction. | Increased punch height, causing a 0.063 in (1.6 mm) gap between the trial and the instrument. This may cause the surgeon to select an insert that is too thin. Surgical delay may result if the punch dislodges from the inserter and needs to … |
| 2013-12-09 | Z-1077-2014 | Class II | DePuy Orthopaedics, Inc. | PFC SIGMA Cruciate Retaining (CR); Cemented Femoral devices; (Size 5/RT) Used during primary total knee arthroplasty to improve patient mobility. | Knee implant femoral devices were found to have anomalous microstructure. Porosity and the presence of blocky carbides, within the cobalt chrome part, can decrease the overall fatigue strength. This decrease in strength could cause implants … |
| 2013-02-05 | Z-0971-2013 | Class II | DePuy Orthopaedics, Inc. | Specialist 2 IM Rod, Non-Sterile and associated Surgical Techniques: Sigma HP Instruments Classic Surgical Technique,0612-89-510 Sigma HP Instruments Balanced Surgical Technique, 0612-88-510 Sigma HP Instruments Fixed Reference Surgical … | Potential for the IM rod to break, leaving fragments in the patient. Surgical Techniques guidance has been updated until design changes can be made to reduce likelihood of potential failure. |
From FDA enforcement reports that name this device, by a company paying for it. Class I is the most serious recall and Class III the least.
The full lists, detailed tables and downloads are in the Data view.