OASIS

Medical device: Animal-derived wound matrix dressing. 4 companies paid 1,890 providers $629K in 2025 payments naming it.

  • Medical device
  • Paid for by Smith+Nephew, Inc., Atrium Medical Corporation +2
  • FDA Class U
Paid to providers, 2025 $629K 3,040 payments to 1,890 providers
Device problem reports 699 1 deaths, 165 injuries (FDA MAUDE)
Companies paying for it 4 in 2025 payments

Companies paying for it (2025)

Safety record

  • FDA recalls naming it13
  • Device problem reports699
  • FDA risk classClass U

Reports to FDA don't prove the product caused the event.

About this device

A bioabsorbable animal-derived protein matrix [e.g., peptides derived from collagen] designed to provide: 1) a matrix of structural proteins and/or 2) a skin template of living cells [fibroblasts], to facilitate the infiltration/attachment of native cells/tissues (e.g., blood vessels) for wound/skin healing. It is available in various forms [e.g., solution, powder, occlusive/semi-occlusive cover (non-oral)] and may include additional compounds (e.g., alginate, fatty acids). It may be intended to be covered with a secondary dressing. It is commonly intended for treating partial- and …

Brand
OASIS
Company
COOK BIOTECH INCORPORATED
Model described
OASIS XL (G46646)
Device type
Animal-derived wound matrix dressing
FDA product code
KGN · Dressing, wound, collagen
FDA class
Class U
Specialty
Unknown
Sold
by prescription
Status
In Commercial Distribution
GUDID record
00827002466463

FDA clearance or approval

Premarket decisions the device cites
NumberKindDeviceApplicantDecided
K061711510(k)OASIS WOUND MATRIXCook Biotech, Inc. 2006-07-19

A 510(k) clearance means FDA found the device much like one already sold. A PMA approval means FDA reviewed clinical evidence, which it requires for high-risk devices.

FDA's device database (GUDID) record of the device identifier named most in the payments: one model of the product line.

FDA recalls

13 recalls naming it since 2004 (Class I, the most serious: 0; Class II: 13; Class III: 0).

The newest FDA recalls naming the product
StartedRecallClassFirmProductReason
2023-11-01Z-0621-2024Class IIAtrium Medical Corporation Atrium Oasis Drain Single W / AC, Part No. 3600-100A sterilization nonconformance prematurely aged the product by 48.3 days, resulting in the expiration date on the label not being accurate. This nonconforming product was inadvertently distributed to customers. Increased risk to patients …
2023-09-18Z-0317-2024Class IIAtrium Medical Corporation Oasis Dry Suction Water Seal Chest Drains (Oasis): (1)3600-100 DRAIN, OASIS SINGLE W/AC; (2)3612-100 DRAIN, OASIS PEDI A/C; (3)3650-100 DRAIN, OASIS BRU W/AC;Instructions for Use (IFU) for the Atrium Ocean, Oasis, and Express chest drains do not provide sufficient precaution instruction for proper set up of catheter(s) and patient tube connections with single collection chamber chest drains
2023-09-05Z-0486-2024Class IIAtrium Medical Corporation Atrium Oasis Dry Suction Water Seal Chest Drain, Part Number: 3650-100, DRAIN, OASIS BRU W/AC, 3650 ATS BLOOD RECOVERYProduct was re-processed and re-sterilized by a third party using packaging configurations and a sterilization process which are not approved by the manufacture and lack validation.
2023-09-05Z-0485-2024Class IIAtrium Medical Corporation Atrium Oasis Dry Suction Water Seal Chest Drain, Part Number: 3600-100, DRAIN, OASIS SINGLE W/AC, 3600 SINGLE COLLECTION.Product was re-processed and re-sterilized by a third party using packaging configurations and a sterilization process which are not approved by the manufacture and lack validation.
2016-12-23Z-1211-2017Class IIAtrium Medical Corporation Maquet Express Dry Seal Suction OASIS DRAIN, DRY PEDI W/AC Reference: 3612-100 Intended Use: To evacuate air and/or fluid from the chest cavity or mediastinum. " To help re-establish lung expansion and restore breathing dynamics. " To …Chest drains manual vent valve located at the top of affected chest drains could potentially not re-seal after being depressed.
2016-12-23Z-1207-2017Class IIAtrium Medical Corporation Maquet Oasis Dry Suction Water Seal Chest Drain, Single Collection w/AC Reference: 3600-100 Intended Use: To evacuate air and/or fluid from the chest cavity or mediastinum. " To help re-establish lung expansion and restore breathing …Chest drains manual vent valve located at the top of affected chest drains could potentially not re-seal after being depressed.
2016-11-29Z-1086-2017Class IIAtrium Medical Corporation Atrium Medical Oasis Blood Recovery Drains as follows: 3650-100 Oasis Blood Recovery drain with in-line connector 3652-100 Oasis Double Blood Recovery drain with in-line connectors - 2 patient lines connected by one Y connector …Outer Packaging is not sterile
2016-11-29Z-1085-2017Class IIAtrium Medical Corporation Atrium Medical 3620-100: Oasis dual chamber drain with in-line connectors (2 Patient Tubes) Product Usage: To evacuate air and/or fluid from the chest cavity or mediastinum.  To help re-establish lung expansion and restore breathing …Outer Packaging is not sterile
2016-11-29Z-1084-2017Class IIAtrium Medical Corporation Atrium Medical Oasis Pediatric Drain as follows: 3612-100 Oasis Infant/Pediatric drain with in-line connector, ¿ in. tubing, Sims and Pedi connector in bag on back of drain 3612-400 Oasis Infant/Pediatric drain with no in-line …Outer Packaging is not sterile
2016-11-29Z-1083-2017Class IIAtrium Medical Corporation Atrium Medical Oasis Single Drain as follows: 3600-100 Oasis single drain with in-line connector 3600-150 Oasis single drain with in-line connector and Pre-connected 2550 ATS Bag Product Usage: To evacuate air and/or fluid from the …Outer Packaging is not sterile
2015-05-07Z-1937-2015Class IIAtrium Medical Corporation Oasis Dry Suction Water Seal Chest Drain, DRY SINGLE W/AC P/N: 3600-100Chest drains have a missing o-ring
2013-11-19Z-0741-2014Class IIAtrium Medical Corporation Atrium Oasis Dry Suction Water Seal ATS (Autotransfusion) Blood Recovery System Chest Drain Dual patient lines; Product Code: 3652-100. To evacuate air and/or fluid from the chest cavity or mediastinum.Chest Drain tubing of of the ATS Blood recovery may leak or disconnect
2013-11-19Z-0740-2014Class IIAtrium Medical Corporation Atrium Oasis Dry Suction Water Seal ATS (Autotransfusion) Blood Recovery System Chest Drain with in-line connectors Product Code: 3650-100. To evacuate air and/or fluid from the chest cavity or mediastinum.Chest Drain tubing of of the ATS Blood recovery may leak or disconnect

From FDA enforcement reports that name this device, by a company paying for it. Class I is the most serious recall and Class III the least.

The full lists, detailed tables and downloads are in the Data view.