HOFFMANN

Medical device: External orthopaedic fixation system noninvasive component, single-use. 1 company paid 1,519 providers $3.00M in 2025 payments naming it.

  • Medical device
  • Paid for by Stryker Corporation
  • FDA Class 2
Paid to providers, 2025 $3.00M 3,708 payments to 1,519 providers
Device problem reports 912 0 deaths, 418 injuries (FDA MAUDE)
Companies paying for it 1 in 2025 payments

Companies paying for it (2025)

Safety record

  • FDA recalls naming it16
  • Device problem reports912
  • FDA risk classClass 2 (moderate)

Reports to FDA don't prove the product caused the event.

About this device

A noninvasive, non-powered structural component of an external orthopaedic fixation system intended for bone/joint treatment (typically fracture stabilization, deformity correction, or distraction), in conjunction with an invasive bone pin (not included). It is in the form of a full or partial ring, rod (e.g., straight, curved, telescopic), plate, or connector (e.g., screw, bolt, nut, hinge, coupling, fastener); it is not a surgical instrument. This is a single-use device.

Brand
HOFFMANN
Company
Stryker GmbH
Model described
Connecting Rod (5028-8-100)
Device type
External orthopaedic fixation system noninvasive component, single-use
FDA product code
LXT · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT, METAL COMPOSITE
FDA class
Class 2 (moderate risk)
Specialty
Orthopedic
Sold
by prescription
Status
In Commercial Distribution
GUDID record
04546540492852

FDA clearance or approval

Premarket decisions the device cites
NumberKindDeviceApplicantDecided
K051306510(k)HOFFMANN II MRI EXTERNAL FIXATION SYSTEMHowmedica Osteonics Corp. 2005-06-17

A 510(k) clearance means FDA found the device much like one already sold. A PMA approval means FDA reviewed clinical evidence, which it requires for high-risk devices.

FDA's device database (GUDID) record of the device identifier named most in the payments: one model of the product line.

FDA recalls

16 recalls naming it since 2004 (Class I, the most serious: 0; Class II: 16; Class III: 0).

The newest FDA recalls naming the product
StartedRecallClassFirmProductReason
2026-07-01Z-2833-2026Class IIStryker GmbH Stryker Hoffmann II Connecting Rod Micro 3 x 150mm. Catalog Number: 50796150.Potential cracking of the Hoffmann II Carbon Connecting Rods following cleaning and sterilization.
2026-07-01Z-2832-2026Class IIStryker GmbH Stryker Hoffmann II Connecting Rod Micro 3 x 120mm. Catalog Number: 50796120.Potential cracking of the Hoffmann II Carbon Connecting Rods following cleaning and sterilization.
2026-07-01Z-2831-2026Class IIStryker GmbH Stryker Hoffmann II Connecting Rod Micro 3 x 90mm. Catalog Number: 50796090.Potential cracking of the Hoffmann II Carbon Connecting Rods following cleaning and sterilization.
2026-07-01Z-2830-2026Class IIStryker GmbH Stryker Hoffmann II Connecting Rod Micro 3 x 60mm. Catalog Number: 50796060.Potential cracking of the Hoffmann II Carbon Connecting Rods following cleaning and sterilization.
2026-07-01Z-2829-2026Class IIStryker GmbH Stryker Hoffmann II Connecting Rod Micro 3 x 50mm. Catalog Number: 50796050.Potential cracking of the Hoffmann II Carbon Connecting Rods following cleaning and sterilization.
2026-07-01Z-2828-2026Class IIStryker GmbH Stryker Hoffmann II Connecting Rod Micro 3 x 40mm. Catalog Number: 50796040.Potential cracking of the Hoffmann II Carbon Connecting Rods following cleaning and sterilization.
2026-07-01Z-2827-2026Class IIStryker GmbH Stryker Hoffmann II Connecting Rod Micro 3 x 30mm. Catalog Number: 50796030.Potential cracking of the Hoffmann II Carbon Connecting Rods following cleaning and sterilization.
2024-05-09Z-2219-2024Class IIStryker GmbH Stryker Hoffman LRF Wire Tensioner REF 4933-9-100, a component of the Hoffmann LRF System.The adjustment ring on the device may become loose resulting in the wire tension not being able to be set correctly.
2023-02-07Z-1191-2023Class IIStryker GmbH stryker Hoffmann LRF, Transport Strut, Model Number 4933-0-400; Orthopedic external fixation device intended to keep bone fragments/segments in a correct anatomical position until reliable bone consolidation is achievedStryker has identified a nonconformance in specific lots of Hoffmann LRF Bone Transport Struts. Specifically, the thread pitch is oversized at the distal end of the threaded rod, and as a result the distal ring fixation cannot be locked …
2019-11-21Z-0781-2020Class IIStryker GmbH Stryker Pin To Rod Coupling External Fixation System Hoffmann II Compact - Pin To Rod Coupling 5/3-4mm Catalog #: 4940-1-020 - Product Usage: Indications Distal radius fractures (intra and extra) articular Intercarpal fracture …Pin To Rod Coupling could not clamp the pin and rod as intended
2018-08-28Z-0291-2019Class IIStryker GmbH Hoffman LRF Hexapod Strut, Extra Short 183mm-295mm, Model Number 4935-0-040. Product Usage - The Hoffmann LRF System is indicated in pediatric patients and adults for the treatment and fixation of: Open and closed fractures …The press fit connection of the Angular Sleeve and pin of the U-Joint and Click Lock did not meet specification. A dimensional discrepancy of the bore hole of the Angular Sleeve may cause a reduced press fit of the pin. This can result in …
2018-08-28Z-0290-2019Class IIStryker GmbH Hoffman LRF Hexapod Strut, Extra Short 131mm-191mm, Model Number 4935-0-030. Product Usage - The Hoffmann LRF System is indicated in pediatric patients and adults for the treatment and fixation of: Open and closed fractures …The press fit connection of the Angular Sleeve and pin of the U-Joint and Click Lock did not meet specification. A dimensional discrepancy of the bore hole of the Angular Sleeve may cause a reduced press fit of the pin. This can result in …
2018-08-28Z-0289-2019Class IIStryker GmbH Hoffman LRF Hexapod Strut, Extra Short 105mm-139mm, Model Number 4935-0-020. Product Usage - The Hoffmann LRF System is indicated in pediatric patients and adults for the treatment and fixation of: Open and closed fractures …The press fit connection of the Angular Sleeve and pin of the U-Joint and Click Lock did not meet specification. A dimensional discrepancy of the bore hole of the Angular Sleeve may cause a reduced press fit of the pin. This can result in …
2018-08-28Z-0288-2019Class IIStryker GmbH Hoffman LRF Hexapod Strut, Extra Short 89mm-109mm, Model Number 4935-0-010. Product Usage - The Hoffmann LRF System is indicated in pediatric patients and adults for the treatment and fixation of: Open and closed fractures …The press fit connection of the Angular Sleeve and pin of the U-Joint and Click Lock did not meet specification. A dimensional discrepancy of the bore hole of the Angular Sleeve may cause a reduced press fit of the pin. This can result in …
2018-07-26Z-3240-2018Class IIStryker GmbH Hoffmann LRF Hexapod Software Catalog #: 983031 Software Version #: HLRF 1.0-2/002, HLRF 1.1-2/002When creating a correction plan with the Correct Axial First button selected and the No. of corrections per day set to more than 1 (e.g., 2, 3, or 4), the Hexapod Software computes an erroneously accelerated correction plan for the axial …
2018-03-12Z-1526-2018Class IIStryker GmbH Hinge Coupling Hoffmann LRF Catalog # 4933-0-800 Intended for fixation of fractures, joint contractures, fusions, limb lengthening, deformity correction, bone and soft tissue reconstruction in pediatric patients and adults.The wrong parts were picked for packaging. The connecting Nuts were packed in the packaging labeled with Hinge coupling

From FDA enforcement reports that name this device, by a company paying for it. Class I is the most serious recall and Class III the least.