GALAXY

Medical device: Interproximal dental gauge, reusable. 1 company paid 308 providers $971K in 2025 payments naming it.

  • Medical device
  • Paid for by Noah Medical Corporation
  • FDA Class 1
Paid to providers, 2025 $971K 1,822 payments to 308 providers
Device problem reports 3,489 32 deaths, 727 injuries (FDA MAUDE)
Companies paying for it 1 in 2025 payments

Companies paying for it (2025)

Safety record

  • FDA recalls naming it5
  • Device problem reports3,489
  • FDA risk classClass 1 (lowest)

Reports to FDA don't prove the product caused the event.

About this device

A device intended to be used in dentistry to measure the clearance between adjacent teeth (e.g., in orthodontic treatment during reproximation to relieve pressure points between teeth). The device is typically available in a set of stainless steel graduated gauges, typically with notches on the side to help with identification. Device thickness typically ranges from 0.1 mm to 0.5 mm. This is a reusable device.

Brand
Galaxy
Company
ORTHO TECHNOLOGY, INC.
Model described
Galaxy Interproximal Reduction 6pc Gauge Set (IPR-GAUGE)
Device type
Interproximal dental gauge, reusable
FDA product code
EIL · GAUGE, DEPTH, INSTRUMENT, DENTAL
FDA class
Class 1 (lowest risk)
Specialty
Dental
Sold
by prescription
Status
In Commercial Distribution
GUDID record
00190730030702

FDA's device database (GUDID) record of the device identifier named most in the payments: one model of the product line.

FDA recalls

5 recalls naming it since 2004 (Class I, the most serious: 0; Class II: 5; Class III: 0).

The newest FDA recalls naming the product
StartedRecallClassFirmProductReason
2025-09-12Z-0598-2026Class IINoah Medical Corporation Brand Name: Galaxy System Product Name: Galaxy System Model/Catalog Number: GAL-001 Product Description: The Galaxy SystemTM and its accessories are intended to provide bronchoscopic visualization of and access to patient airways for …Due to Users not having access to the Instructions for Use/User Manual due to it being password protected.
2024-10-28Z-0742-2025Class IINoah Medical Galaxy System (GAL-001). Intended to provide bronchoscopic visualization of and access to patient airways for diagnostic and therapeutic procedures.Bronchoscopic visualization system power supplies have defective semiconductor components that may fail unexpectedly, which could result in an aborted procedure since it is a non recoverable failure. Manual removal of the bronchoscope when …
2024-10-25Z-0693-2025Class IINoah Medical Galaxy System (GAL-001)Bronchoscopic system uninterruptable power may have defective component, may cause power supply short circuit/trip protective circuit breakers causing sudden, unrecoverable power loss, which could occur while system is on mains or internal …
2024-10-25Z-0679-2025Class IINoah Medical Galaxy bronchoscope (Product GALB-001) used with Instructions for use, Part: 10000752, part of the Noah Medical Galaxy System (GAL-001)Not all biopsy tools meant to fit bronchoscope channel (2.1mm inner diameter/length greater or equal to 1100mm) will pass through if scope channel positioned in sharp bend (approx. greater than 100-deg.), excessive force may result in …
2024-03-09Z-1734-2024Class IINoah Medical Galaxy System REF GAL-001 (specifically GAL-019, Wired Controller) The Galaxy System and its accessories are intended to provide bronchoscopic visualization of and access to patient airways for diagnostic and therapeutic procedures.Due to a misaligned calibration of the wired controller that may result in unintended motion of the scope.

From FDA enforcement reports that name this device, by a company paying for it. Class I is the most serious recall and Class III the least.

The full lists, detailed tables and downloads are in the Data view.