DJO SURGICAL
Medical device: Total reverse shoulder prosthesis. 1 company paid 479 providers $29.5M in 2025 payments naming it.
- Medical device
- Paid for by Encore Medical, Lp
- FDA Class 2
Companies paying for it (2025)
- ENCORE MEDICAL, LP$29.5M
Safety record
- FDA recalls naming it16
- Device problem reports5,240
- FDA risk classClass 2 (moderate)
Reports to FDA don't prove the product caused the event.
About this device
A sterile implantable artificial substitute for a shoulder joint designed to replace both articulating surfaces of the joint particularly in patients with chronic, longstanding rotator cuff muscle tears, arthritis, and/or humeral fractures unsuccessfully treated by previous surgery. The device reverses normal anatomy so that the head portion of the joint is attached to the scapula and the socket portion is attached to the humerus. It typically includes a humeral stem and glenosphere (head) with body and cup components. The device may include screws for implantation and may be implanted with or …
- Brand
- DJO SURGICAL
- Company
- Encore Medical, L.P.
- Model described
- Retaining Screw, RSP 8mm Spacer (510-08-001)
- Device type
- Total reverse shoulder prosthesis
- FDA product code
- KWS · Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented
- FDA class
- Class 2 (moderate risk)
- Specialty
- Orthopedic
- Sold
- —
- Implant
- Yes
- Status
- In Commercial Distribution
- GUDID record
- 00888912024860
FDA clearance or approval
| Number | Kind | Device | Applicant | Decided |
|---|---|---|---|---|
| K111061 | 510(k) | REVERSE SHOULDER MONOBLOCK PROSTHESIS (RSP) | Encore Medical L.P. | 2011-07-20 |
A 510(k) clearance means FDA found the device much like one already sold. A PMA approval means FDA reviewed clinical evidence, which it requires for high-risk devices.
FDA's device database (GUDID) record of the device identifier named most in the payments: one model of the product line.
FDA recalls
16 recalls naming it since 2004 (Class I, the most serious: 0; Class II: 16; Class III: 0).
| Started | Recall | Class | Firm | Product | Reason |
|---|---|---|---|---|---|
| 2025-07-18 | Z-2549-2025 | Class II | Encore Medical, LP | Brand Name: DJO SURGICAL Product Name: FA S Altivate Reverse Glenoid Tray Model/Catalog Number: None Orthopedic instrument for shoulder prosthesis | The reamer may kick or bind up during or immediately prior to use. |
| 2025-07-18 | Z-2548-2025 | Class II | Encore Medical, LP | Brand Name: DJO SURGICAL Product Name: ALTIVATE REVERSE WEDGE GLENOID REAMER SLEEVE Model/Catalog Number: 804-06-310 Product Description: Material: S.S/CoCrMo, Non-Sterile Orthopedic instrument for shoulder prosthesis | The reamer may kick or bind up during or immediately prior to use. |
| 2025-07-18 | Z-2547-2025 | Class II | Encore Medical, LP | Brand Name: DJO SURGICAL Product Name: ALTIVATE REVERSE WEDGE GLENOID REAMER BOSS DRILL, 6.5mm Model/Catalog Number: 804-06-312 Product Description: Material: S.S., Non Sterile Orthopedic instrument for shoulder prosthesis | The reamer may kick or bind up during or immediately prior to use. |
| 2025-07-18 | Z-2546-2025 | Class II | Encore Medical, LP | Brand Name: DJO SURGICAL Product Name: ALTIVATE REVERSE WEDGE GLENOID REAMER HEAD Model/Catalog Number: 804-06-311 Product Description: Material: S.S/Nitronic, Non-Sterile Orthopedic instrument for shoulder prosthesis | The reamer may kick or bind up during or immediately prior to use. |
| 2023-11-10 | Z-0508-2024 | Class II | Encore Medical, LP | DJO surgical EMPOWR Porous Knee System Empowr 3D Femur-Porous Coated SIZE 8 Left 3D MATRIX COATED REF 243-01-108 DJO surgical EMPOWR KNEE", PRESS FIT BP MINUS, 8L REF 353-03-108 For the treatment of patients who are candidates for … | Due to incorrect product/device within packaging. |
| 2023-09-06 | Z-0532-2024 | Class II | Encore Medical, LP | RSP Glenoid Baseplate, 30mm P2 Porous Coated REF 508-32-204. DJO Surgical shoulder device. | Coating hex did not engage with the Straight Mod Hexdriver or with the Hex Head Screw Driver |
| 2023-08-04 | Z-2607-2023 | Class II | Encore Medical, LP | DJO Surgical: Altivate Reverse, Humeral Stem, Small Shell, Size 6 x 108mm, REF: 533-06-108 | Orthopedic device components were not subjected to final sterilization before shipment, which if implanted may lead to varying degrees of infection, which may require treatment, prolonged monitoring, and may lead to revision surgery |
| 2023-08-04 | Z-2606-2023 | Class II | Encore Medical, LP | DJO Surgical: Reverse Shoulder System (RSP) Semi Humeral Socket Insert, 32mm, REF: 508-01-032; Baseplate, Glenoid HA-Coat, RSP, 6.5mm X 30mm, REF: 508-32-104; Reverse Shoulder System (RSP) Glenoid, Neutral Head W/ Retaining Screw, 36mm, … | Orthopedic device components were not subjected to final sterilization before shipment, which if implanted may lead to varying degrees of infection, which may require treatment, prolonged monitoring, and may lead to revision surgery |
| 2023-08-04 | Z-2605-2023 | Class II | Encore Medical, LP | DJO Surgical: CLP Hip, Stem, Polished Neck, Lateral Offset Implant, Size#3, REF: 426-11-030 | Orthopedic device components were not subjected to final sterilization before shipment, which if implanted may lead to varying degrees of infection, which may require treatment, prolonged monitoring, and may lead to revision surgery |
| 2023-08-04 | Z-2604-2023 | Class II | Encore Medical, LP | DJO Surgical: Delta Ceramic Femoral Head, 44 mm + 4.0 mm, REF: 400-03-443 | Orthopedic device components were not subjected to final sterilization before shipment, which if implanted may lead to varying degrees of infection, which may require treatment, prolonged monitoring, and may lead to revision surgery |
| 2023-08-04 | Z-2603-2023 | Class II | Encore Medical, LP | DJO Surgical: Empowr Knee, Universal Tibia Baseplate, 6R, REF: 352-02-106 | Orthopedic device components were not subjected to final sterilization before shipment, which if implanted may lead to varying degrees of infection, which may require treatment, prolonged monitoring, and may lead to revision surgery |
| 2021-05-12 | Z-2104-2021 | Class II | Encore Medical, LP | DJO Surgical EMPOWR 3D Knee Tibial Insert - e-plus, REF: 341-16-711 Left Sz. 11 16mm, CE 0086, Sterile H2O2, UDI: (01)00888912167185 | Packaging mixup; There is a potential that packaging identified as 12mm knee prosthesis contains 16mm knee prosthesis and 16mm knee prosthesis packaging contains 12 mm knee prosthesis. |
| 2021-05-12 | Z-2103-2021 | Class II | Encore Medical, LP | DJO Surgical EMPOWR 3D Knee Tibial Insert - e-plus, REF: 341-12-711, Left Sz. 11 12mm, CE0086, Sterile H2O2, UDI: (01)00888912166980 | Packaging mixup; There is a potential that packaging identified as 12mm knee prosthesis contains 16mm knee prosthesis and 16mm knee prosthesis packaging contains 12 mm knee prosthesis. |
| 2020-08-31 | Z-0117-2021 | Class II | Encore Medical, LP | AltiVate Reverse Small Socket Insert, 44mm Semi Constrained - E-Plus, Part Number: 509-03-044 - Product Usage: Per IFU, DJO Surgical shoulder devices are intended for treatment of patients who are candidates for shoulder arthroplasty per … | Product mislabeled. Use may result in a delay in surgery, stopping of surgery, and revision surgery. |
| 2018-09-21 | Z-1123-2019 | Class II | Encore Medical, Lp | Exprt Precision System: Revision Hip, Proximal Body with Bolt, Lateral Offset, Sterile R, djo surgical, REF: 495-00-065, 495-00-075, 495-00-085, 495-01-065, 495-01-075, 495-01-085 | Complaints regarding loose proximal body bolts or bolts that were found through x-ray as sitting above the proximal body and not properly seated. Patient risks may include: pain, dissociation, and possible revision surgery |
| 2012-07-16 | Z-2146-2012 | Class II | Encore Medical, Lp | djo surgical Screw 6.5 mm, Sz 25mm, low profile. Product is intended for the fixation of the acetabular shell or fracture repair | One lot of the 010-55-030, 6.5 mm Cancellous Bone Screw was recalled because a product complaint identified a mislabeling of the device. Although the package label indicates Size 30mm, the screw dimension is actually 25mm. |
From FDA enforcement reports that name this device, by a company paying for it. Class I is the most serious recall and Class III the least.
The full lists, detailed tables and downloads are in the Data view.