Constellation
Medical device: Phacoemulsification/vitrectomy system generator. 2 companies paid 277 providers $407K in 2025 payments naming it.
- Medical device
- Paid for by Alcon Vision Llc, Alcon Research Llc
- FDA Class 2
Companies paying for it (2025)
- ALCON VISION LLC$348K
- ALCON RESEARCH LLC$59K
Safety record
- FDA recalls naming it17
- Device problem reports27,713
- FDA risk classClass 2 (moderate)
Reports to FDA don't prove the product caused the event.
About this device
A mains electricity (AC-powered) ophthalmic device, that is a component of a system, that generates the energies (e.g., ultrasonic, piezoelectric, high frequency current) delivered through handpieces/tips that are introduced through an incision(s) made in the eye, for the purposes of performing combined phacoemulsification and/or vitrectomy (the removal and replacement of cataracts and/or the vitreous in anterior and posterior segment surgical procedures). It will typically be activated by the surgeon using a foot-switch and may include an irrigation/suction system for the removal of debris, a …
- Brand
- CONSTELLATION
- Company
- Alcon Laboratories, Inc.
- Model described
- CONSTELLATION TABLE TOP (8065753041)
- Device type
- Phacoemulsification/vitrectomy system generator
- FDA product code
- HQE · Instrument, vitreous aspiration and cutting, ac-powered
- FDA class
- Class 2 (moderate risk)
- Specialty
- Ophthalmic
- Sold
- by prescription
- Status
- In Commercial Distribution
- GUDID record
- 00380657530410
FDA's device database (GUDID) record of the device identifier named most in the payments: one model of the product line.
FDA recalls
17 recalls naming it since 2004 (Class I, the most serious: 0; Class II: 16; Class III: 1).
| Started | Recall | Class | Firm | Product | Reason |
|---|---|---|---|---|---|
| 2025-09-15 | Z-0581-2026 | Class II | Alcon Research LLC | CONSTELLATION ULTRAVIT 10K and CONSTELLATION HYPERVIT 20K, Description/REF: 25+ TOTALPLUS CP PAK 20K CPM BWV .9 IU/ 8065000093; 25+ TOTALPLUS CP PAK 20K CPM BV .9 IU/ 8065000095; 27+ TOTALPLUS CP PAK 20K CPM BV .9 IU/ 8065000096; 23GA … | XXX |
| 2025-08-05 | Z-0308-2026 | Class II | Alcon Research LLC | Brand Name: Alcon Laboratories, Inc. Product Name: Accessory devices for Unity and Constellation surgical consoles Model/Catalog Number: Many - See 'Attachment 1 - Affected Product List' 8065000356 27GA HP ENTRY SYSTEM, 4 MM 1 COUNT … | Due to incomplete seals in the pouch which provide the sterile barrier. |
| 2025-08-05 | Z-0307-2026 | Class II | Alcon Research LLC | Brand Name: Alcon Laboratories, Inc. Product Name: Accessory devices for Unity and Constellation surgical consoles Model/Catalog Number: Many - See 'Attachment 1 - Affected Product List' 8065000252 UNITY 27 GA CHANDELIER … | Due to incomplete seals in the pouch which provide the sterile barrier. |
| 2025-08-05 | Z-0306-2026 | Class II | Alcon Research LLC | Brand Name: Alcon Laboratories, Inc. Product Name: Accessory devices for Unity and Constellation surgical consoles Model/Catalog Number: Many - See 'Attachment 1 - Affected Product List' 8065000393 UNITY HP VISCOUS FLUID CONTROL VFC PACK … | Due to incomplete seals in the pouch which provide the sterile barrier. |
| 2025-08-05 | Z-0305-2026 | Class II | Alcon Research LLC | Brand Name: Alcon Laboratories, Inc. Product Name: Accessory devices for Unity and Constellation surgical consoles Model/Catalog Number: Many - See 'Attachment 1 - Affected Product List' 8065750918 PAK,I/A LINE,NGVS … | Due to incomplete seals in the pouch which provide the sterile barrier. |
| 2025-08-05 | Z-0304-2026 | Class II | Alcon Research LLC | Brand Name: Alcon Laboratories, Inc. Product Name: Accessory devices for Unity and Constellation surgical consoles Model/Catalog Number: Many - See 'Attachment 1 - Affected Product List' 8065000250 UNITY 27+DS HYPERVIT 30K VIT PROBE … | Due to incomplete seals in the pouch which provide the sterile barrier. |
| 2025-04-28 | Z-2109-2025 | Class II | Alcon Research LLC | Brand Name: Alcon Laboratories, Inc. Product Name: Constellation Vision System Combined Procedure Pak Model/Catalog Number: Many - See 'Attachment 1 - Affected Product List' Software Version: N/A Product Description: Surgical procedure … | Due to the potential that some units within specific lots were damaged during the manufacturing process. The area of damage is within the same area of the lidding. While the damage in the lidding is visibly noticeable, due to risk of … |
| 2025-04-28 | Z-2108-2025 | Class II | Alcon Research LLC | Brand Name: Alcon Laboratories, Inc. Product Name: Constellation Vision System Vitrectomy Procedure Pak Model/Catalog Number: Many - See 'Attachment 1 - Affected Product List' Software Version: N/A Product Description: Surgical … | Due to the potential that some units within specific lots were damaged during the manufacturing process. The area of damage is within the same area of the lidding. While the damage in the lidding is visibly noticeable, due to risk of … |
| 2024-10-15 | Z-0490-2025 | Class III | Alcon Research LLC | a) Alcon Custom-Pak CONSTELLATION 25G PLUS CHRISTIANA CARE HLTH SERV, # 15340-11, containing TUBING, SALEM SUMP, 14 FR; b) Alcon Custom-Pak DR GREEN VITRECTOMY MERCY HOSPITAL, # 6557-59 and # 6557-60 each containing TUBING, SALEM SUMP, … | Reported breakage in the tubing Anti-Reflux Valve (ARV) due to improper use. As a result, changes to labeling were made. |
| 2024-07-15 | Z-2647-2024 | Class II | Alcon Research LLC | Product Name: 25+ TTL PLUS VPAK 20000CPM BWV Model/Catalog Number: 8065830077 Software Version: N/A Product Description: Standardized assemblage of components, which includes the valved trocar entry system, that interface with the … | Alcon has detected an increase in complaint reports related to sharpness for specific ClearCut Dual Bevel Sideport and A-OK Comeal/Scleral V-Lance Ophthalmic knives. |
| 2023-08-18 | Z-0099-2024 | Class II | Alcon Research, LLC | Constellation Vision System, REF: 8065751150 | Ophthalmic microsurgical vision systems have a new Printed Circuit Board (PCB) design that could lead to higher voltage to a PCB component, that could cause component failure and loss of pressure control for the use of cutters, scissors, … |
| 2023-06-28 | Z-2406-2023 | Class II | Alcon Research, LTD. | Constellation Procedure Pak | Reports of patient injury during use in procedures due to potential for plastic debris from over-tightening with included plastic wrench. |
| 2018-07-25 | Z-1794-2020 | Class II | Alcon Research LLC | Alcon Constellation Vision System 20 GA TOTALPLUS Combined Procedure Pak, Product #8065751079, Sterile. The firm name on the label is Alcon Laboratories, Inc., Fort Worth, TX. | There was a potential that a 25-gauge infusion cannula may be included within the procedure pak instead of a 20-gauge infusion cannula. |
| 2018-01-12 | Z-0779-2018 | Class II | Alcon Research, LTD. | Custom Pak¿ (Custom Pak in which Alcon 25+¿ TOTALPLUS¿ Vitrectomy Pak is a component.) Includes all necessary items for a core vitrectomy procedure and connects to the CONSTELLATION Vision System. | Alcon is conducting this Voluntary Medical Device Recall for a specific lot (2060953H) of its 25+ TOTALPLUS Vitrectomy Pak due a potential that a 23GA ULTRAVIT probe may be included within the 25+ TOTALPLUS Vitrectomy Pak, instead of a 25+ … |
| 2018-01-12 | Z-0778-2018 | Class II | Alcon Research, LTD. | Alcon 25+¿ TOT ALPLUS¿ Vitrectomv Pak Includes all necessary items for a core vitrectomy procedure and connects to the CONSTELLATION Vision System. | Alcon is conducting this Voluntary Medical Device Recall for a specific lot (2060953H) of its 25+ TOTALPLUS Vitrectomy Pak due a potential that a 23GA ULTRAVIT probe may be included within the 25+ TOTALPLUS Vitrectomy Pak, instead of a 25+ … |
| 2015-08-11 | Z-1693-2016 | Class II | Alcon Research, Ltd. | 23G and 25G Constellation Totalplus Paks provides access into the posterior segment of the eye for surgical instruments within the frame of 23G and 25G Pars-Plana-Vitrectomy (PPV) surgeries. | The product has potential to leak beyond their design specification. |
| 2012-05-29 | Z-2470-2012 | Class II | Alcon Research, LTD. | This voluntary medical device correction is related to Alcon products that present a 23G nonvalved trocar cannula and infusion cannula together. These items are available as components within Accurus 23G TotalPLUS Pak, Accurus 23G Combined … | This medical device correction is related to Alcon products that present a 23G non-valved trocar cannula and infusion cannula together because of an issue regarding the loose fit between the 23G non-valved trocar cannula and infusion … |
From FDA enforcement reports that name this device, by a company paying for it. Class I is the most serious recall and Class III the least.
The full lists, detailed tables and downloads are in the Data view.