APTUS

Medical device: Orthopaedic bone screw, non-bioabsorbable, non-sterile. 1 company paid 290 providers $761K in 2025 payments naming it.

  • Medical device
  • Paid for by Medartis Inc.
  • FDA Class 2
Paid to providers, 2025 $761K 442 payments to 290 providers
Device problem reports 65 1 deaths, 36 injuries (FDA MAUDE)
Companies paying for it 1 in 2025 payments

Companies paying for it (2025)

Safety record

  • FDA recalls naming it6
  • Device problem reports65
  • FDA risk classClass 2 (moderate)

Reports to FDA don't prove the product caused the event.

About this device

A small, non-sterile, threaded rod with a slotted head typically used for internal orthopaedic fracture fixation by being screwed into bone to hold plates or nails to bone or to provide direct interfragmentary stabilization of bone, or it may fasten soft tissue to bone; it is made of a material that cannot be chemically degraded or absorbed via natural body processes (includes implant grade metal such as surgical steel, titanium alloy, or carbon fibre). Screws available are: cortical, cancellous, malleolar, scaphoid, partially-threaded and fully-threaded, self-tapping, and cannulated. A "lag …

Brand
APTUS
Company
Medartis AG
Model described
2.8 Cortical Screw 40mm, HD7, 1/Pkg (A-5800.40/1)
Device type
Orthopaedic bone screw, non-bioabsorbable, non-sterile
FDA product code
HWC · Screw, fixation, bone
FDA class
Class 2 (moderate risk)
Specialty
Orthopedic
Sold
by prescription
Status
In Commercial Distribution
GUDID record
07630008427207

FDA clearance or approval

Premarket decisions the device cites
NumberKindDeviceApplicantDecided
K091479510(k)APTUS FOOT SYSTEMMedartis AG 2009-07-13

A 510(k) clearance means FDA found the device much like one already sold. A PMA approval means FDA reviewed clinical evidence, which it requires for high-risk devices.

FDA's device database (GUDID) record of the device identifier named most in the payments: one model of the product line.

FDA recalls

6 recalls naming it since 2004 (Class I, the most serious: 0; Class II: 6; Class III: 0).

The newest FDA recalls naming the product
StartedRecallClassFirmProductReason
2026-10-20Z-0433-2027Class IIMedartis AG APTUS 1.2/1.5 TriLock Scaph.Pl 3x2H Large,t0.8, Model/Catalog Number: A-4350.81; APTUS fixation systems are intended for temporary fixation, correction or stabilization of bonesPlate is incorrectly bent
2026-04-24Z-2361-2026Class IIMedartis AG Medartis APTUS, 5.0 headedCCS Long Thread 50/20mm, REF A-8216.50/1, 1/Pkg; Cannulated Compression Screws for bone fixationThe cannulation of the CCS screw is not centered.
2026-01-09Z-1550-2026Class IIMedartis AG Brand Name: APTUS Product Name: 2.8 TriLock Screw 16mm, HD7, 1/Pkg Model/Catalog Number: A-5850.16/1 Software Version: Not applicable. Product Description: The APTUS fixation systems are used for fractures, osteotomies and arthrodesis …Mix up between specified lots of 2.5mm and 2.8mm outer diameter screws.
2026-01-09Z-1549-2026Class IIMedartis AG Brand Name: APTUS Product Name: 2.5 TriLock Screw 16mm, HD7, 1/Pkg Model/Catalog Number: A-5750.16/1 Software Version: Not applicable. Product Description: The APTUS fixation systems are used for fractures, osteotomies and arthrodesis …Mix up between specified lots of 2.5mm and 2.8mm outer diameter screws.
2024-04-12Z-1821-2024Class IIMedartis Inc. APTUS 2.8 TriLock Screw 22mm, HD7, 1/Pkg. Metal bone fixation fastener used with the APTUS fixation system.Screw is 16mm long instead of 22mm long.
2014-10-01Z-0150-2015Class IIMedartis, Inc. APTUS Ulna Shortening 2.5 Product Usage: APTUS Ulna Shortening 2.5 Plates are indicated for fractures and osteotomies, in particular for the ulna.A recall of the APTUS Ulna Shortening 2.5 instrument was initiated due to three (3) complaints of breakage of the temporary tension bolt.

From FDA enforcement reports that name this device, by a company paying for it. Class I is the most serious recall and Class III the least.

The full lists, detailed tables and downloads are in the Data view.