ALINITY
Medical device: Chemiluminescent immunoassay analyser IVD. 1 company paid 253 providers $151K in 2025 payments naming it.
- Medical device
- Paid for by Abbott Laboratories
- FDA Class 1
Companies paying for it (2025)
- ABBOTT LABORATORIES$151K
Safety record
- FDA recalls naming it14
- Device problem reports11,319
- FDA risk classClass 1 (lowest)
Reports to FDA don't prove the product caused the event.
About this device
An electrically-powered automated or semi-automated laboratory instrument intended to be used for the qualitative and/or quantitative in vitro determination of chemical and/or biological markers (e.g., protein, drug, hormone, microbial toxin) in a clinical specimen, using an immunological method that utilizes a chemiluminescent detection system together with sample processing, data processing and/or data display software.
- Brand
- Alinity
- Company
- ABBOTT LABORATORIES
- Model described
- Alinity i Processing Module (03R6501)
- Device type
- Chemiluminescent immunoassay analyser IVD
- FDA product code
- JJE · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE
- FDA class
- Class 1 (lowest risk)
- Specialty
- Clinical Chemistry
- Sold
- —
- Status
- In Commercial Distribution
- GUDID record
- 00380740137366
FDA's device database (GUDID) record of the device identifier named most in the payments: one model of the product line.
FDA recalls
14 recalls naming it since 2004 (Class I, the most serious: 0; Class II: 12; Class III: 2).
| Started | Recall | Class | Firm | Product | Reason |
|---|---|---|---|---|---|
| 2025-11-13 | Z-0961-2026 | Class II | Abbott Laboratories | Alinity ci-series System Control Module (SCM), REF: 03R70-01, 03R70-20 | Abbott Laboratories is recalling their Alinity ci-series System Control Module, a chemistry/immunoassay anayzer, by correction. The reason for the recall is potential performance issues found in the Alinity ci-series System software … |
| 2025-09-18 | Z-0489-2026 | Class II | Abbott Laboratories | Brand Name: Alinity hq Analyzer Product Name: Alinity hq Analyzer Model/Catalog Number: 09P68-01 Software Version: 5.0 to 5.10.1 for OUS, 5.7 and 5.10.1 in the US Product Description: Alinity hq Analyzer Component: No | Software issue for hq analyzer results in system not visibly applying appropriate flagging to results when saturation is present which may result in incorrect results. |
| 2024-06-14 | Z-2681-2024 | Class II | Abbott Laboratories | Alinity hq Analyzer, REF: 09P68-01 | 1)If Complete Blood Count with differential and reticulocyte(CBC+Diff+Retic) sample is run, then immediately following on same rack CBC+Diff sample is run, then falsely low red blood cell count may occur, generating falsely high Mean Cell … |
| 2023-10-19 | Z-0681-2024 | Class II | Abbott Laboratories | Alinity c Processing Module REF 03R67-01 The Alinity c Processing Module is a fully automated chemistry analyzer allowing random and continuous access, as well as priority and automated retest processing using photometric and … | The devices contain a dry natural rubber (latex) subcomponent, which are not labeled as containing dry natural rubber(latex). |
| 2023-10-19 | Z-0680-2024 | Class II | Abbott Laboratories | Alinity i Processing Module REF 03R65-01 The Alinity i Processing Module is a fully automated immunoassay analyzer allowing random and continuous access, as well as priority and automated retest processing using chemiluminescent … | The devices contain a dry natural rubber (latex) subcomponent, which are not labeled as containing dry natural rubber(latex). |
| 2023-05-30 | Z-0613-2024 | Class II | Abbott Laboratories | Alinity ci-series System Control Module, REF: 03R70-01 | There are potential performance issues found in the Alinity ci-series System software versions 3.4.0 and lower, including: 1) SCC Reagent load error during daily maintenance; 2) Sample Laboratory Report error; 3) Customer requests mechanism … |
| 2022-11-24 | Z-0958-2023 | Class III | Abbott Laboratories | Alinity Reaction Vessels, REF: 06P1401, for use with Alinity i Systems; ARCHITECT Reaction Vessels | Reaction vessels may contain a potential contaminant that could affect their optical performance. |
| 2022-06-09 | Z-1456-2022 | Class II | Abbott Laboratories | REF: 08P4377, Alinity c Hemoglobin Reagent Kit, IVD, CE, R1 5 x 46.0 mL, R2 5 x 18.3 mL, A1cDL 5 x 67.6 mL | There is a potential for falsely elevated Hemoglobin %A1c and Hemoglobin A1c results to be generated using the ARCHITECT and Alinity c Hemoglobin A1c assay. |
| 2022-06-09 | Z-1455-2022 | Class II | Abbott Laboratories | REF: 08P4320, Alinity c Hemoglobin Reagent Kit, IVD, CE, R1 5 x 46.0 mL, R2 5 x 18.3 mL, A1cDL 5 x 67.6 mL | There is a potential for falsely elevated Hemoglobin %A1c and Hemoglobin A1c results to be generated using the ARCHITECT and Alinity c Hemoglobin A1c assay. |
| 2022-02-03 | Z-0749-2022 | Class II | Abbott Laboratories, Inc | Alinity s System software version2.8.0, REF LN 06P16-01, IVD, OEI: (01)00380740138479 The Alinity s System is intended for In Vitro diagnostic use only. The Alinity s System is a highthroughput, fully automated immunoassay analyzer … | Software error associated with the immunoassay analyzer wash cycle which is using 1 mL of wash buffer instead of the intended 3 mL of wash buffer to wash the exterior of the probe. |
| 2021-07-26 | Z-2472-2021 | Class II | Abbott Laboratories | Alinity ci-series System Control Module clinical chemistry and immunoassay analyzer for in vitro diagnostic use. Model number LN 03R70-01 | Due to potential performance issues with software version 3.2.3 and earlier. Performance issues are: 1) There is the potential for onboard reagents past their lot expiration date or onboard stability time to be incorrectly displayed on … |
| 2021-07-12 | Z-2276-2021 | Class II | Abbott Laboratories, Inc | Alinity s System REF 06P16-01. An automated immunoassay analyzer for In Vitro diagnostic use in blood donor screening, plasma and plasmapheresis screening, and organ donor centers, hospitals, and reference laboratories. | A design defect (hardware and software) allows liquid waste pressure to build up and potentially spray users/operators. |
| 2021-06-10 | Z-2184-2021 | Class II | Abbott Laboratories | REF: 08P1820, Abbott Alinity c Activated Alanine Aminotransferase reagent Kit, Alinity c Activated Alanine Aminotransferase Reagent Kit,, IVD, CE, UDI: (01) 00380740132569 | There is a potential for 15% under-recovery on Activated Alanine Aminotransferase (A-ALT) samples greater than 1,200 U/L. In addition there is a potential for a >10% shift in patient results with samples containing bilirubin concentrations … |
| 2019-04-09 | Z-1361-2019 | Class III | Abbott Laboratories | ALINITY i Anti-TPO Calibrators - Product Usage: The Alinity i Anti-TPO Calibrators are for the calibration of the Alinity i analyzer when used for the quantitative determination of the IgG class of thyroid peroxidase autoantibodies … | Devices were delivered without the required dry ice. |
From FDA enforcement reports that name this device, by a company paying for it. Class I is the most serious recall and Class III the least.
The full lists, detailed tables and downloads are in the Data view.