ABRE
Medical device: Bare-metal peripheral venous stent. 1 company paid 1,117 providers $235K in 2025 payments naming it.
- Medical device
- Paid for by Medtronic, Inc.
- FDA Class 3
Companies paying for it (2025)
- MEDTRONIC, INC.$235K
Safety record
- FDA recalls naming it3
- Device problem reports326
- FDA risk classClass 3 (highest)
Reports to FDA don't prove the product caused the event.
About this device
A non-bioabsorbable, tubular device made of metal intended to be implanted in a peripheral (typically the iliac and/or femoral) vein to maintain patency in patients with constrained venous outflow; it is not intended for coronary or intracranial venous stenting. It is inserted and advanced to the site of implantation with a dedicated catheter where it self-expands upon release. It is typically made of nickel-titanium alloy (Nitinol) and has radiopaque marker(s) for visualization; it is available in a variety of diameters and lengths. Disposable devices associated with implantation may be …
- Brand
- Abre™
- Company
- MEDTRONIC, INC.
- Model described
- STENT AB9U16120090 ABRE V01 (AB9U16120090)
- Device type
- Bare-metal peripheral venous stent
- FDA product code
- QAN · Stent, iliac vein
- FDA class
- Class 3 (highest risk)
- Specialty
- Unknown
- Sold
- by prescription
- Implant
- Yes
- Status
- In Commercial Distribution
- GUDID record
- 00643169796362
FDA clearance or approval
| Number | Kind | Device | Applicant | Decided |
|---|---|---|---|---|
| P200026 | PMA | Abre Venous Self-expanding Stent System | Medtronic Vascular, Inc. | 2020-10-21 |
A 510(k) clearance means FDA found the device much like one already sold. A PMA approval means FDA reviewed clinical evidence, which it requires for high-risk devices.
FDA's device database (GUDID) record of the device identifier named most in the payments: one model of the product line.
FDA recalls
3 recalls naming it since 2004 (Class I, the most serious: 0; Class II: 3; Class III: 0).
| Started | Recall | Class | Firm | Product | Reason |
|---|---|---|---|---|---|
| 2025-12-04 | Z-1098-2026 | Class II | Medtronic Inc | Abre Venous Self-Expanding Stent System, Model numbers AB9G14120090 AB9G12100090 AB9G14080090 AB9G12080090 AB9U12060090 AB9U14100090 AB9U14080090 AB9U12080090 AB9U14060090 AB9G14100090 AB9G12060090 | Stents may be impacted by a Nitinol anomaly which causes a reduced stent diameter when deployed and corresponding reduced radial force. Issue may result in stent migration. |
| 2024-02-16 | Z-1317-2024 | Class II | Medtronic Inc. | Abre Venous Self-expanding Stent System, Product Numbers: a) AB9U14080090; b) AB9U18100090; c) AB9G14100090 | There is a potential for sterile package breach. |
| 2021-03-16 | Z-1532-2021 | Class II | Medtronic Inc. | Medtronic Abre Venous Self-expanding Sent System, Catalog number REF AB9U16080090 - Product Usage: intended for use in the iliofemoral veins for the treatment of symptomatic venous outflow obstruction. | Incorrect size printed on the device; packaging is labeled correctly. |
From FDA enforcement reports that name this device, by a company paying for it. Class I is the most serious recall and Class III the least.
The full lists, detailed tables and downloads are in the Data view.